[Home] [Catalog] [Category] [Previous Item] [Next Item]

Calibrators

Calibrators Digital Venturi Flow - 110 V

[Item Image]

D-814 Venturi Flow Calibrator
Precision digital microprocessor controlled
air flow calibrator utilizing a venturi flow sen-
sor. Air flow range is 2 to 14 CFM, 56 to 400
LPM @ STP. UL, CSA
D-814
***********************************************************************************************
************

MODEL D-814 DIGITAL VENTURI CALIBRATOR

***********************************************************************************************
************

____________________________________________________________________
_________

Certifications
____________________________________________________________________
_________

- UL 3111-1 Standards for Electrical Measuring and Test Equipment 1st Ed.

- CSA-C22.2 No. 1010-1-92 Safety Requirements for Electrical Equipment for
Measurement
Control and Laboratory Use Part 1.

____________________________________________________________________
________

Notable Features
____________________________________________________________________
_________

- Manufactured under an ISO 9001 quality assurance program.
- Flow measuring device: Venturi.
- Display of flow in CFM, LPM or cubic meters per minute by operator selection.
- Flow corrected to standard temperature and pressure (STP).
- Display of barometric pressure in metric or English units.
- Display of temperature in metric or English units.
- Portable battery operated unit is available.
- Optional STP flow and actual flow model is available. Refer to Model D-814AS.
- Standard Accuracy: +/- 2.0% of full scale.
- Optional Accuracy: +/- 1.0% of full scale. Refer to Model D-814-1.

____________________________________________________________________
_________

Specifications - Measurement Principles
____________________________________________________________________
_________

Flow Sensor Venturi tube.

Standardization Correction for standard temperature and
barometric
pressure.

Curve Linearization Individually calibrated and software corrected.

Measurement Ranges 2 to 14 CFM, 56.0 to 400.0 LPM.

Temperature 40-104 degrees F (4-40
degree C)
Barometric Pressure 30-24 in of Hg (760-
609 mm Hg)

Measurement Accuracy Air Flow +/- 2.0% of full scale
(Optional +/- 1.0% of full
scale)
Temperature +/- 0.6 degree F @ 25
degree C)
Typical = 0.9 degree C
at 0 degree
and 100
degree C
Maximum error = +/-1.5
degree C
+/-
2.7 degree F
Barometric Pressure +/- 1% over
measurement range

Displayed Parameters and Resolution
Display is 4 digit LCD 0.7" high.

Parameter Engineering Unit
Resolution
CFM cubic feet per minute 0.01
CFM
LPM liter per minute 0.1
LPM
m^3/min cubic meter per minute 0.01
m^3/min
degree F degree Fahrenheit 0.1
degree F
degree C degree Celsius 0.1
degree C
In-Hg inches of Mercury 0.01 In
-Hg
mm-Hg millimeters of Mercury mm-Hg

____________________________________________________________________
_________

Specifications - General:
____________________________________________________________________
_________

Air Flow Range 2 to 14 CFM, 56.0 to 400.0 LPM.

Power Requirements 100-120 VAC; 50/60 Hz.

Operating Temperature 40-104 degree F; (4-40 degree C).

Storage Temperature: 32-158 degrees F; (0-70 degree C).

Dimensions 9 3/4" L x 8 1/4" W x 10 3/4" H.
(250 mm x 212 mm x 275 mm).

Weight 8 pounds 6 ounces; (3.75 kg).

Calibration Factory calibration is recommended once per
year.

____________________________________________________________________
_________

Options
____________________________________________________________________
_________

Actual flow/STP flow switch to display actual flow or STP flow.

RS232 interface, including cable.
All measured and calculated parameters are accessible through RS232 port to a
computer.

Data acquisition and processing software is available upon request.

Portable battery option. Refer to Model D-814B.
- The battery pack is inside the instrument case.
- The unit will operate on AC power if it is available. It also charges the battery.
The
calibrator operates from the internal battery source when AC power is not
available.
- The battery pack consists of 7.2 V, 2.2 Ah size batteries, 6 each of C size
batteries (1.2v)
in series. Battery type, NiCad.
- A separate battery supervisor circuit is responsible for charging the battery,
switching over
from AC to battery, the full charge indicator and the low battery detector.
- The operating time on the battery is greater than 6 hours.
- The recharge time is 8 hours.
- The battery life is 2 years.

____________________________________________________________________
_________

Venturi Tube
____________________________________________________________________
_________

Correction for standard temperature and barometric pressure. Individually
calibrated and
software corrected.

Air Flow Range
Type CFM LPM
V-814 2 to 14 56.0 to 400.0


32 to 104 degree F (0 to 40 degree C)
30 to 24 inches Hg (760 to 609 mm Hg)
Approximate sea level to 6000 foot elevation above sea level.

F&J SPECIALTY PRODUCTS, INC.

PO Box 2888
Ocala, FL 34478-2888
404 Cypress Road
Ocala, FL 34472
Telephone: (352) 680-1177
Fax: (352) 680-1454