Home : Air Flow Calibrators : Digital Air Flow Calibrators


DIGITAL FLOW CALIBRATORS for AIR SAMPLERS
Regular Line Power and with Battery Powered Options
 

NOTABLE FEATURES:
  • Flow measuring device: Venturi
  • Display of flow in CFM, LPM or M3/min. by operator selection (LED display)
  • Standard accuracy: ±2% Full Scale
  • Flowrates displayed are corrected to a factory settable Reference Temperature and Pressure (4 options available)
    Classical STP 0° C 1 Atm
    Normal T and P 20° C 1 Atm
    Modified Normal T and P 70° F 1 Atm
    Standard Ambient T and P 25° C 1 Atm
  • Display of barometric pressure in metric or English units
  • Display of temperature in metric or English units
  • Portable battery operated unit is available
  • Certified to UL, CSA and CE electrical safety standards by ITSNA (formerly ETL or Edison Testing Laboratories)
  • Ambient or Reference flow is selectable by the operator
  • 2 year warranty

Email or Contact Us for pricing or information!

MODEL SELECTION


Standard Units Battery Option Units Flow Range
100-120V 230V 100-120V 230V CFM LPM
D-801V.2 D-801EV.2 D-802BV.2 D-802BEV.2 0.1 - 1 3 - 30
D-802V.2 D-802EV.2 D-802BV.2 D-802BEV.2 0.2 - 2 6 - 56
D-812V.2 D-812EV.2 D-812BV.2 D-812BEV.2 0.5 - 4 14 - 115
D-828V.2 D-828EV.2 D-828BV.2 D-828BEV.2 1 - 9 28-255
D-814V.2 D-814EV.2 D-814BV.2 D-814BEV.2 2 - 14 56 - 400
D-530V.2 D-530EV.2 D-530BV.2 D-530BEV.2 5 - 30 140 - 850
D-540V.2 D-540EV.2 D-540BV.2 D-540BEV.2 5 - 40 140 - 1130
D-870V.2 D-870EV.2 D-870BV.2 D-870BEV.2 8 - 70 230-1980

Options:

  • Battery Powered Operation (Denoted by ‘B’ in model number)
  • 1% F.S. accuracy units available (add ‘-1’ to model number); example: D-812E-1
  • Actual Flow/STP Flow Display (Denoted by A/S in model number)
  1. MEASUREMENT PRINCIPLE
    1.1 Flow Sensor: Venturi Tube
    1.2 Standardization: Correction for standard temperature and barometric pressure
    1.3 Curve linearization: Individually calibrated and software corrected

  2. MEASUREMENT RANGES
    2.1 Air flow: See Table on front page
    2.2 Temperature: -4° F - +122°F (-20°C – +50° C)
    2.3 Barometric pressure: 30 – 24 In-Hg (760 – 609 mm-Hg)
    approx. Sea level to 6000 ft. elevation above sea level

  3. MEASUREMENT ACCURACY
    3.1 Air flow: +/- 2% of full scale (Optional +/- 1% of full scale)
    3.2 Temperature (Typical): 1°C over range -4° F - +122°F (-20°C – +50° C)
    3.3 Barometric pressure: +/- 1% over measurement range

  4. DISPLAY 6 Digit LCD 0.4" high

  5. DISPLAYED PARAMETERS and RESOLUTIONS

      Parameter Engineering Unit

    Resolution

    5.1 CFM Cubic feet per minute 0.01 CFM
    5.2 LPM Liter per minute 0.1 or 1 LPM*
    5.3 M3/m Cubic meter per minute 0.0001 M3/min
    5.4 F Degree Fahrenheit 0.1 Degree F
    5.5 C Degree Celsius 0.1 Degree C
    5.6 In-Hg Inches of Mercury 0.01 In-Hg
    5.7 mm-Hg Millimeters of Mercury 1 mm-Hg

    * A 0.1 LPM resolution applies to calibrators that have a maximum flow of 900 LPM
     

  6. GENERAL

    6.1 Power requirements: 100–120 VAC;50/60Hz 0.3 Amps max; 220-240 VAC/50Hz
    6.2 Operating temperature: -4° F - +122°F (-20°C – +50° C)
    6.3 Storage temperature: -20° F - +122°F -29°C – +70° C)
    6.4 Dimension (L×W×H) 9.75"×8.25"×10.75" (248×210×273mm)
    6.5 Weight: 8 lbs. 6 oz. (3.8 Kgs.)
    6.6 Installation Category: Pollution Degree 2  
    6.7 Enclosure Rating: IPXO  

  7. CALIBRATION
    Factory calibration is recommended once per year
     
  8. OPTIONS
    8.1 RS-232 interface, including cable — All measured and calculated parameters are
    accessible through an RS-232 port to a computer.
    8.2 Data acquisition and processing software is available upon request.
    8.3 AC input of 230V, 50Hz is available upon request at no additional charge.

F&J SPECIALTY PRODUCTS, INC.
404 Cypress Road
Ocala, FL 34472
Tel: 352-680-1177
Fax: 352-680-1454
Email: fandj@fjspecialty.com