Home : Tritium Collection Systems : TCS-3000

TRITIUM COLLECTION SYSTEM
F&J MODEL TCS-3000

NOTABLE FEATURES:
  • RS-232 Port for real-time data
  • Large, bright LED Display
  • Microprocessor controlled electronics
  • Flowrate measurements and volume totalizations at actual or STP conditions
  • Precision machined DP sensor
  • Flowrate accuracy within ± 4% F.S.
  • Flowrate / Volume Options:
    sccm / scc
    SLPM / SL
  • LED dimming feature
  • 110-120VAC,50/60Hz, single phase
  • 220V Systems available

Email or Contact Us for pricing or information!

  


GENERAL DESCRIPTION:

The Model TCS-3000 Tritium Collection System is a tritium collection system consisting of a diaphragm pump, airflow regulating rheostat and 2 removable polycarbonate silica gel columns. The flow and volume of air passing through the system is measured by a microprocessor controlled Digital Flowmeter (DFM). The DFM utilizes a precision-machined differential pressure sensor to measure flowrate. The DFM displays on-board calculations on a bright large character LED display. Flowrate and totalized volume corrected to STP and elapsed time are displayed.

Multiple operator selectable data download frequencies are available through the RS232 port for collection and/or storage of real-time data.

The unit is designed for continuous indoor use. Please consult the product specifications for the design temperature range and the installation category.

The typical operating flow range is 30 - 300 sccm (0.03 – 0.3 LPM).

TCS-3000 Tritium Collection System (100 – 120VAC)


SPECIFICATIONS:

PUMP TYPE: Diaphragm

MAXIMUM CAPACITY: Minimum 50 cc/min; Maximum 400 cc/min.

POWER REQUIREMENTS: 100 – 120VAC; 50/60 Hz; 2 ampere; single phase

CIRCUIT BREAKER PROTECTION: 5 amperes

ELECTRICAL CORD: All temperature, 3-wire, 16 gauge

DIMENSIONS: 9"D × 24"W × 20"H

WEIGHT: 67 lbs. (30.3 Kgs.)

INSTALLATION CATEGORY: Pollution Degree 2

ELECTRONIC SPECIFICATIONS

MEASUREMENT ACCURACY

Air flow: ± 4% of full scale

Temperature: ± 0.9ºF (0.5ºC) (Not displayed)

Barometric Pressure: ± 0.6 inches Hg (Not displayed)

OPERATING TEMPERATURES: 0º-104ºF (-17º-40ºC)

STORAGE TEMPERATURE: -20º-122ºF (-28º-50ºC)

CALIBRATION: Factory calibration once per year

COMMUNICATIONS INTERFACES: RS-232 available for realtime data download.

 

ON-BOARD CALCULATIONS

  • Flow calculation from differential pressure value corrected to STP
  • Elapsed Time
  • Cumulative Volume corrected to STP

OPTIONS:

  • FlashCard Datalogger system for collection and storage of real-time data exiting the RS232 port.
  • FlashCard data storage device: P/N: 232FCDSD
  • FlashCard 128 MB; P/N: 372239
  • FlashCard Reader; P/N: 515177

NOTE:

Other tritium absorbing media may be utilized, such as molecular sieve, water or ethylene glycol.

 


F&J SPECIALTY PRODUCTS, INC.
404 Cypress Road
Ocala, FL 34472
Tel: 352-680-1177
Fax: 352-680-1454